3411 Torque Display Firmware Update |
|
The 3411 firmware version F0 (and before) is not capable of being updated in the field. Only if you have version G0 are you capable of using this feature. If you would like to upgrade your 3411 unit to firmware version G0, please contact your Magtrol sales representative. |
Version AA0 - Released 03/2017 |
|
|
Changes made this version: |
- Low frequency filters were incorrectly reporting torque due to floating point round off error. The variables used to calculate torque have been changed from floats to doubles removing the round-off error produced during the calculations.
|
|
Version V0 - Released 05/2016 |
|
|
Changes made this version: |
- Speeds below 5 Hz are correctly reported in Torque 7 and the “OB” command as 0 Hz.
USB reading at 500 Hz is now stable when using Ethernet communication simultaneously.
|
|
Version U0 - Released 01/2016 |
|
Changes made this version: |
- Raw speed reported correctly when changes are made to pulse per rev value
|
|
Version T0 - Released 12/2015 |
|
|
Changes made this version: |
- Speed is displayed correctly when changes are made to the encoder pulse per rev setup option
|
|
Version R0 - Released 11/2015 |
|
|
Changes made this version: |
- Analog I/O now outputs correct voltage.
- Speed readings are clamped at 5 Hz, input signals at lower frequencies are now set to a speed of 0
|
|
|
|
|
|